MaxQ 2000 CO2 Plus Digital CO2 Resistant Shaker Catalog #88881101, #88881102 #88881103, #88881104 Operation Manual 0000847 In the United States: In Canada: For customer service, call 1-800-766-7000 For customer service, call 1-800-234-7437 To fax an order, use 1-800-926-1166 To fax an order, use 1-800-463-2996 To order online: thermofisher.com...
Page 2
We can also provide you with a quotation on our Extended Warranty for your Thermo Scientific products. Whatever Thermo Scientific products you need or use, we will be happy to discuss your applications. If you are experiencing technical problems, working together, we will help you locate the problem and, chances are, correct it yourself...over the...
Flask Clamp Platform & Starter Kit Flask Clamp 500ml Starter Kit Flask Clamp Main Body Flask Clamp Control Box Flask Clamp General Power Adaptor US Plug CN Plug EU Plug UK Plug Section 1 Unpacking and Installation MaxQ 2000 CO2 Plus...
2) Control box 3) Main body of CO2 resistant shaker the main body, please cut off the power recessed countersink head screws. 3) Power adapter first. ▲ 4) Power cord Section 1 | Unpacking and Installation MaxQ 2000 CO2 Plus...
2.2 Environmental Conditions Application Environmental Conditions: indoor use Temperature..............4 to 60℃ Altitude................≤2,000 m Humidity..............20% to 99% Concentration............0% to 20% Storage Environmental Conditions Temperature..............5 to 60℃ Altitude................≤2,000 m Humidity........20% to 99%, non-condensing Section 2 | Overview MaxQ 2000 CO2 Plus...
▲ ≤500ml 500ml 2) Please refer to the "Liquid Volume" table in the specifications section for the ≤500ml 1000ml maximum load. ▲ ≤1000ml ≤1000ml 2000ml ≤2000ml ≤1500ml 3000ml ≤3000ml ≤2500ml 5000ml ≤5000ml Section 2 | Overview MaxQ 2000 CO2 Plus...
1. When the shaker is running, the LED rapidly. display will switch between the speed Key: The power is On/Off. and timer function every 30 seconds Key: The instrument is running or stopped. Section 3 | Operation MaxQ 2000 CO2 Plus...
3. When the main body of CO2 Resistant or on the shaker. Shaker is in operation in the CO2 incubator, keep a safe distance from the inner wall of the CO2 incubator to avoid collisions. Section 4 | Application MaxQ 2000 CO2 Plus...
Section 7 Troubleshooting To troubleshoot the unit, refer to the contact your local sales/service table below. If your problem persists, representative. Error Cause Solution Connect the power Power disconnected Indicator off Power adapter failure Replace power adapter No reaction Switch on Switched off Sensor failure Contact local sales/service...