Section 7
Initial Operation
System Cleaning
Procedure (cont.)
7-6
Nanopure Life Science (UV/UF)
16. Press the START/STOP switch to return the unit to the normal
recirculation mode.
17. Remove the new 0.2 micron filter and bell assembly from the bag and
insert it into the dispenser arm manifold. Gently turn it clockwise until
it is fully seated.
18. Remove the protective cap from the filter bell. Turn the dispense knob
to the left or right and flush 1-2 liters of water through the 0.2 micron
filter. The product water is now ready for use.
Figure 7-3. Installing 0.2 Micron Filter
Note For more demanding applications where low TOC water is required,
a rinse of 15-20 liters through the cartridges and 0.2 micron filter may be
necessary.
s
Caution Never dispense water without first installing hose barb provided in
parts bag or final filter supplied with cartridge pack.