Section 5
Installation

UV Lamp Installation

Other Accessories

5-2
Nanopure TOC - UV
Locate the UV chamber inside the left door of the Nanopure TOC - UV.
Install the UV Lamp as follows:
1. Disconnect the power cord from the unit.
2. Access the left side of the unit by
unlatching the screw securing the
door with a flathead screwdriver.
Open the door. The UV lamp
will be installed vertically.
3. Remove the UV lamp from its
packaging. DO NOT TOUCH
THE GLASS PORTION OF
THE LAMP! It is recommended
that lint-free gloves be worn when
handling the lamp. The glass
portion must be free of
fingerprints, perspiration, etc.
Even a light coating of
perspiration will reduce the
effectiveness of the lamp. If the
glass portion of the lamp is
touched, clean it with a damp,
lint-free cloth: use isopropyl alcohol as required.
4. Remove black cap by sliding off UV housing.
5. Insert the lamp halfway into the UV lamp chamber and plug it in,
then fully insert the lamp into the chamber.
6. Make sure the UV lamp cable is tucked behind the chamber before
closing and relatching the door.
Note The UV lamp contains mercury. If broken or no longer needed, do
not dispose of the UV lamp in the trash. Recycle or dispose of the UV
lamp as hazardous waste.
Optional accessories for this unit include a float switch, low pressure
switch, N.I.S.T. Traceable Calibration Module, standard remote dispenser,
Accudispense volumetric remote dispenser, dispense overflow cutoff float
and printer. If you purchased a float switch or low pressure switch, refer to
"User Settings' in this manual. If you purchased a standard or
Accudispense volumetric remote dispenser, refer to their respective operator
manuals for installation instructions.
Figure 5-1. UV Lamp Installation -