Commissioning - Smeg BPW1260 User Manual

Bedpan washers – hospital sector
Hide thumbs Also See for BPW1260:
Table of Contents

Advertisement

8 COMMISSIONING

Once the device has been installed correctly, it must be prepared for use.
Open the bottom compartment and switch the device on using the main switch (position shown in the
illustration below).
The following operations must be carried out:
1. Set the current date and time (see sections 10.3, 10.4).
2. Set the language required for the printer (optional) (see section 10.2).
3. Set up dosing of the descaler and any detergent used (see section 8.1).
4. On first installation:
a. the peristaltic pumps must be activated by hand by the authorised engineer to ensure that
the pipes in the circuit are primed correctly (see section 8.1).
b. Run Service cycle "Pr 10".
5. Run a program complete with thermal disinfection with no load for processing in the tank (see
sections 9.5, 9.6).
6. The authorised engineer can proceed with complete verification and recording of correct operation
on the user's premises, accordance with the Manufacturer's procedure (ref. SR152).
The device is now ready for use.
User Manual
193909105 rev.02
BPW1260
Page 37 – 93