Smeg BPW1260 User Manual page 48

Bedpan washers – hospital sector
Hide thumbs Also See for BPW1260:
Table of Contents

Advertisement

The cycle can be interrupted in advance at the end of the thermal disinfection phase (once the A
for the cycle has been reached) to avoid having to wait for the cooling and hydraulic circuit pumpout phases
to be completed. To do this, press the "padlock" button (P7).
Once thermal disinfection has finished (from the start of the cooling phase, identified by "END" and "COOL"
alternating on the display) the door can be opened. There are 2 possible scenarios:
Case 1) Door opened with cooling (COOL) still in progress:
END – COOL are shown flashing
If the chamber temperature is higher than the temperature set for opening of the door, the chamber
temperature and the
unlock until button P5 is pressed.
The cooling phase is not interrupted during these operations, until the door is opened.
Once the door is opened, the device must complete pumpout of the tank, so the
flash on the display.
Once pumpout of the tank is complete,
During this phase, all buttons are disabled and the "start/stop" icon remains off.
Case 2) Door opened with COOL phase completed and tank pumpout in progress:
END on display (not flashing) – Door open button can be pressed with no need to confirm.
Once the door is open:
-
the program number remains on the top line (e.g. "Pr 03")
-
the word END disappears from the second line, replaced by the information for the set program.
The
icon remains on (tank pumpout in progress).
Once pumpout of the tank is complete,
During this phase, all buttons are disabled and the "Start / Stop" icon remains off.
User Manual
symbol flash on the second line instead of "END" – "COOL": the door does not
goes out and the start/stop icon illuminates
goes out and the start/stop icon illuminates.
193909105 rev.02
BPW1260
value set
0
icon continues to
Page 48 – 93