Smeg BPW1260 User Manual page 55

Bedpan washers – hospital sector
Hide thumbs Also See for BPW1260:
Table of Contents

Advertisement

SETUP menu items
Typical screens (guideline illustrations)
TU – Temperature units
Default °C
Possible values °C / °F
The value chosen sets the way in which
temperatures are
- shown
- printed
- stored in the device's memory
It can only be modified by the authorised engineer.
RC – Repeat cycle: parameter which sets the cycle
repetitions.
Press P2 or P8: "00" can be edited, it starts to flash.
Set value "01", "02" etc. by pressing P2 or P8; the
value continues to flash until P5 is pressed to
confirm.
["00", repetition mode not active, no repetitions
performed.
"01" or above, repetition mode active.
Value = 1, 2, n, cycle performed "n" times as set in
parameter – max value : 50.]
When a value other than "00" is confirmed with P5,
cycle repetition icon "F2" illuminates.
It stays lit even after setup mode is exited to
indicate that this specific function is active.
N.B.:
if the value is "00": only the engineer can modify the
value.
If the value is different from "00": the superuser can
access the edit function and change the value from
"01" up to "nn".
If the value is "01", icon F2 is lit but the program is
only performed once.
Dry – Dryer on/off: parameter which enables
presence/operation of the dryer:
EVLP2 – Two bedpan valve enabling: parameter
that enables use of the valve for processing the
bedpan 2 (EVLPD2).
This parameter can be modified by the superuser.
Default "ON".
If "OF" all phases involving EVLPD2 are bypassed
User Manual
193909105 rev.02
BPW1260
Page 55 – 93