13.1.1 WARNING TABLE

All the codes below refer to warnings; they are not alarms which stop the cycle. Use of the device can be continued,
but repeated or significant warnings
ID
W 41
P1 detergent intake anomaly
W 42
P2 acid intake anomaly
P1 anomaly, pump active when not
W 46
required.
P2 anomaly, pump active when not
W 47
required
Anomaly in "load cooling" phase.
Cooling is included at the end of
the drying phase to bring the load
W 67
processed and the heating
elements to a "safe" temperature
and for satisfactory steam removal
W P1
P1 jerry can empty
W P2
P2 jerry can empty
Internal memory full.
More than 80% of the memory
W 91
capacity available for storing cycle
data has been used.
Drying filter maintenance.
This alarm does not appear on the
W 92
Display but is recorded in the
device memory; the relative LED
illuminates on the display.
Device maintenance.
This alarm does not appear on the
W 93
Display but is recorded in the
device memory; the relative LED
illuminates on the display.
User Manual
should be reported to the after-sales service.
DESCRIPTION
193909105 rev.02
USER ACTION
1. Check that the detergent intake nozzle is correctly
positioned in the can.
2. Check for detergent leaks (pools of detergent around the
device).
Check that the detergents have not passed their use-by date
and are not crystallised. Possible chemical dispensing
problems.
Inspect the load carefully to ensure that it is clean at the end of
the cycle. Report the problem to the after-sales service.
Contact the after-sales service to have maintenance performed
on the product.
Check that there is detergent in the jerry can or that the level
sensor is operating correctly.
Replace the empty jerry can with a new one of the same
detergent, following the instructions in the manual.
If the setup overwrite function is not set as on, the device
informs the user of the % of memory capacity used.
An AF is triggered when the 95% limit is reached: do not
perform any more cycles until the memory has been cleared.
Contact the after-sales service to have the filter changed.
Contact the after-sales service to have maintenance performed
on the product.