5.2

GENERAL WARNINGS

MAXIMUM LOAD WEIGHT
Leaning or sitting on the open door of the device might cause it to tip over, putting people in
danger.
The
The maximum weight which can be loaded on the door, including the weight of the washing
support, must never exceed 10 kg.
MALFUNCTIONS and PERFORMANCE ANOMALIES
In case of operating anomalies or unforeseen events: read the "ALARMS" section of this manual
and, if necessary, contact your nearest Authorised Service Centre.
In the event of water leaks: disconnect the device from the electricity supply and turn off the
water taps.
If the device's interface display remains off or unresponsive even with the main switch correctly
set ON: do not touch the display as this may trigger unexpected operation of parts of the
device. Switch the device off using the main switch. Wait at least 10 seconds after switching off
before switching back on again.
Refer to the "TROUBLESHOOTING" section.
If changes in the device's performance are noticed check:
1. That all operating conditions are as specified by the manufacturer in this manual and
the documentation supplied with the product.
2. That all maintenance procedures and recommended intervals have been complied with,
3. Perform or request a performance qualification to check whether the device's
performance is within the acceptability parameter range (see ISO 15883 standard,
relevant edition).
If the device's performance is non-compliant, contact the authorised after-sales service to have
it checked and repaired (ref. req. MDR 23.4 s, Annex I).
Maintenance: change drying filter (if installed)
The device provides an alert when the filter needs changing. Keeping the filter in operation after
the end of its useful life will lead to poor steam removal results. If the filter is fouled, the air flow-
rate entering the chamber is reduced.
The filter's useful life may be less than the preset nominal time, depending on the amount of dirt
in the environment. If steam removal performance deteriorates, users should have the filter
changed earlier. The device does not signal this type of wear, but only wear arising from the
hours actually worked.
Routine maintenance
The device provides an alert when routine maintenance is required.
These are inspection and maintenance operations which must be performed by authorised staff
to keep the device safe and in good working order.
These maintenance operations are not covered by the product Warranty, which does not include