NOTE When disinfecting unit surfaces always disconnect
the X-ray unit from mains.
All removable patient support parts (bite pieces, chin rest,
chin support, chin cup and temple supports) can be
autoclaved up to 135°C or cleaned with alcohol-based
solutions.
The patient support handles can be cleaned with alcohol-
based solutions.
Other unit surfaces, including the control panel display,
can be cleaned with a soft cloth damped in a mild
cleaning solution.
NOTE Do not use cleaning agents in aerosol or spray form
directly on unit surfaces.
To guarantee user and patient safety and to ensure image
quality the unit must be checked and recalibrated by a
qualified Planmeca service technician once a year or after
every 10 000 exposures if this is sooner. Refer to the
Planmeca ProOne Technical Manual for complete
servicing information.