PANORAMIC EXPOSURE
72
ProMax X-ray unit with Dimax3
NOTE
Maintain audio and visual contact with the patient
and unit during exposure. If the C-arm stops
moving during exposure, release the exposure
button immediately.
You can now drive the C-arm to the entry position by
touching the Go Entry 1 or Go Entry 2 field in the display.
NOTE
If the "Auto return" function is switched on, the C-
arm will automatically return to the ready position
at the end of the exposure. Refer to section 8.6.1
"User preference settings" on page 51 for more
information.
NOTE
If a large number of exposures is taken in rapid
succession the X-ray tube may overheat and a
cooling time will flash in the display (e.g. 00:42).
The cooling time indicates the delay before the
next exposure can be taken.
When you have taken the exposure the image will be
shown on the computer screen. Note that you must
accept the image by clicking the "OK" button - only
accepted images will be stored in the database. Refer
to the Dimaxis User's Manual for further information.