6 Using K7DD Parameters
V
thd .{
Set the alarm threshold (warning) for the voltage total harmonic
distortion of the measurement target.
The alarm will be turned ON if the voltage total harmonic distortion
exceeds this threshold (warning).
I
rms .[
Set the alarm threshold (critical) for the R.M.S. current of the
measurement target.
The alarm will be turned ON if the R.M.S. current exceeds this
threshold (critical).
I
rms .{
Set the alarm threshold (warning) for the R.M.S. current of the
measurement target.
The alarm will be turned ON if the R.M.S. current exceeds this
threshold (warning).
6 - 16
Voltage Total Harmonic Distortion Warning Threshold
Setting-related page
4-2 Feature Measurement (P. 4-3)
R.M.S. Current Critical Threshold
Setting-related page
4-2 Feature Measurement (P. 4-3)
4-3 Simplified Threshold Settings (P. 4-4)
R.M.S. Current Warning Threshold
Setting-related page
4-2 Feature Measurement (P. 4-3)
4-3 Simplified Threshold Settings (P. 4-4)
Range
---
0.0 to 100.0 (%)
Range
5 A
0.000 to 5.500 (A)
25 A
0.0 to 27.5 (A)
100 A
0.0 to 110.0 (A)
200 A
0.0 to 220.0 (A)
400 A
0.0 to 440.0 (A)
Range
5 A
0.000 to 5.500 (A)
25 A
0.0 to 27.5 (A)
100 A
0.0 to 110.0 (A)
200 A
0.0 to 220.0 (A)
400 A
0.0 to 440.0 (A)