5.2.6
Checking sensor coefficients
In order to check the sensor ID and coefficients on the DTI, press the
softkey <MENU/Enter> 4 times.
If Callendar van Dusen coefficients is used for the sensor, the
following message will be shown in the display:
Use the softkey <Reset/SELECT> to scroll through the coefficients:
(example with both sensors using default coefficients)
SENSOR 1: 9402244
R0
:
1.0000000E+02
│
SENSOR 1: 9402244
A
:
3.9080200E-03
│
SENSOR 1: 9402244
B
: -5.8020000E-07
│
SENSOR 1: 9402244
C
: -4.2735000E-12
124745 04
SENSOR 1: 9402244
R0
:
1.0000000E+02
2017-08-10
SENSOR 2: 9402246
R0 :
1.0000000E+02
│
SENSOR 2: 9402246
A
:
3.9080200E-03
│
SENSOR 2: 9402246
B
: -5.8020000E-07
│
SENSOR 2: 9402246
C
: -4.2735000E-12
23