Page 1
Translation of the original operating manual 11/2017 Rev. 1.7.1 of 08/12/2022 Sigma 4-5KRL;Refrigerated Centrifuge for Robot Use;Sigma 4-5KRL IVD; ; Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Sigma 4-5KRL IVD from serial no. 165039 Refrigerated Centrifuge for Robot Use Operating Manual...
Page 2
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb 2 / 113 Translation of the original operating manual,...
Page 4
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb 4 / 113 Translation of the original operating manual,...
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD Table of contents General information ......................9 Importance of the operating manual .................. 9 Intended use ........................9 Warranty and liability......................9 Copyright ........................10 Standards and regulations ....................
Page 6
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD Table of contents Storage and transport ....................... 29 Dimensions and weight ....................29 Storage conditions ......................29 Notes on transport ......................29 Packaging ........................30 Transport safety device ....................
Page 7
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD Table of contents Malfunctions and error correction ..................51 General malfunctions ...................... 51 7.1.1 Emergency lid release ....................52 Table of error codes ......................53 Service contact .......................
Page 8
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD Table of contents Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb 8 / 113 Translation of the original operating manual,...
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 1 General information General information Importance of the operating manual A fundamental requirement for the safe and trouble-free operation of the centrifuge is to be familiar with the fundamental safety instructions and all possible hazards.
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 1 General information → Copyright The copyright concerning the operating manual remains with Sigma Laborzentrifugen GmbH. The operating manual is solely intended for the operator and their personnel.
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 2 Layout and mode of operation → Layout and mode of operation → Layout of the centrifuge 2.1.1 Functional and operating elements Centrifuge lid User interface (see chapter 6.3.1 -...
Page 12
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 2 Layout and mode of operation Serial interface (see chapter 2.1.3 - "Serial interface (pin assignment)") Equipotential bonding screw Fuse Mains power input Fig. 3: Detailed view A - connections Version 11/2017, Rev.
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 2 Layout and mode of operation 2.1.2 Name plate Manufacturer Power consumption Max. speed Max. kinetic energy Serial number Part number Nominal voltage Type Consult operating...
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 2 Layout and mode of operation 2.1.3 Serial interface (pin assignment) There is a floating, normally closed contact between pin 1 and pin 4. When the centrifuge is started, the contact opens, thereby signalling that the centrifuge is ready for operation.
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 2 Layout and mode of operation → Mode of operation → 2.2.1 Centrifugation principle Centrifugation is a process for the separation of heterogeneous mixtures of substances (suspensions, emulsions, or gas mixtures) into their components.
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 2 Layout and mode of operation → 2.2.2.1 Speed, radius, and relative centrifugal force The acceleration g, which the samples are subject to, can be increased by increasing the radius in the rotor chamber and by increasing the speed.
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 3 Safety → Safety → Marking of the unit The following symbols are used on this centrifuge: Arrow indicating the direction On (Power) of rotation...
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 3 Safety → Explanation of the symbols and notes In this operating manual, the following names and symbols to indicate hazards are used: This symbol stands for a direct hazard to the life and health of persons.
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 3 Safety → Responsibility of the operator The operator undertakes to authorise only trained, specialised personnel to work on the centrifuge (see chapter 3.4 - "Requirements concerning the personnel").
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 3 Safety → Requirements concerning the personnel Risk of injury if the personnel are not sufficiently qualified If unqualified personnel perform work on the centrifuge or are present in the danger zone of the centrifuge, hazards result that can cause serious injuries and considerable damage to property.
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 3 Safety The operating personnel must ensure that any serious event which either directly or indirectly had, could have had or could have one of the following...
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 3 Safety → Safety instructions 3.6.1 Electrical safety As protection against electric shock, the centrifuge is equipped with an earthed mains power cable and connector. To ensure the effectiveness of this safety feature, the following must be ensured: •...
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 3 Safety • Do not spin any substances that could damage the material of the rotors and buckets of the centrifuge in any way. Highly corrosive substances, for example, damage the material and affect the mechanical strength of the rotors and buckets.
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 3 Safety → 3.6.5 Chemical and biological safety If pathogenic, toxic, or radioactive samples are intended to be used in the centrifuge, it is in the responsibility of the user to ensure that all necessary safety regulations, guidelines, precautions, and practices are adhered to accordingly.
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 3 Safety 3.6.7 Resistance of plastics Chemical influences have a strong effect on the polymeric chains of plastics, and, therefore, on their physical properties. Plastic parts can be damaged if solvents, acids, or alkaline solutions are used.
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 3 Safety 3.6.9 Service life (for centrifuges according to the IVD regulation) Centrifuge The service life of the centrifuge depends on several factors, e.g. the type and frequency of use, the area of application and the performance of service and maintenance tasks.
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 3 Safety → Safety devices → 3.7.1 Lid lock device The centrifuge can only be started when the lid is properly closed. The electrical lock must be locked. The lid can only be opened when the rotor has stopped.
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 3 Safety → Measures in the event of hazards and accidents • If an emergency arises, switch off the centrifuge immediately! • If in doubt, call the emergency doctor! DANGER →...
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 4 Storage and transport → Storage and transport → Dimensions and weight Sigma 4-5KRL, Sigma 4-5KRL IVD Height: 406 mm Height with open lid: 890 mm...
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 4 Storage and transport Packaging The centrifuge is packaged in a wooden crate. • After taking off the lid, remove the side panels. • Remove the packaging material.
CAUTION Sigma centrifuges are units of protection class I. The centrifuges of this model series have a three-wire power cord with an IEC C13 connector. The removable power cord must not be longer than 3 m!
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 5 Set-up and connection The centrifuges are equipped with a mains power switch with an integrated thermal circuit breaker. • Switch the unit off by actuating the mains power switch.
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 6 Using the centrifuge → Using the centrifuge → Initial start-up • Before the initial start-up, please ensure that your centrifuge is properly set up and installed (see chapter 5 - "Set-up and connection").
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 6 Using the centrifuge → 6.2.2 Installation of rotors and accessories → 6.2.2.1 Installation of the rotor • Open the centrifuge lid by pressing the lid key.
Page 35
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 6 Using the centrifuge Centrifugation with vessels of various sizes Working with vessel of various sizes is possible. In this case, however, it is very important that the rotationally symmetrical inserts are identical (see figure).
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 6 Using the centrifuge 6.2.2.3 Adapters In order to ensure easy handling, even if vessels of various sizes are used, carrier systems were developed. •...
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 6 Using the centrifuge → Control system "Spincontrol L" → 6.3.1 User interface The centrifuge is operated via three buttons with integrated light-emitting diodes and one function knob. The display is divided into several different fields.
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 6 Using the centrifuge → 6.3.2 Manual mode → 6.3.2.1 Starting a centrifugation run The centrifuge is ready for operation when the start key is illuminated.
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 6 Using the centrifuge → 6.3.2.5 Speed / relative centrifugal force (RCF) The set speed of the centrifuge is displayed in the upper area of the Speed/RCF field.
Page 40
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 6 Using the centrifuge Runtime as of the set speed If the runtime is to be counted as of the moment when the set speed is...
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 6 Using the centrifuge Short run A short run can be started if no run is active. • Keep the start key pressed during the short run.
Page 42
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 6 Using the centrifuge Program "Rapid Temp" The centrifuge has a special "Rapid Temp" program that precools the centrifuge rapidly under defined conditions. • Select the program with the aid of the menu item "run Prog" (see chapter 6.3.3.2 - "Loading a program").
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 6 Using the centrifuge → → 6.3.2.8 Rotor selection This field shows the rotor that is currently being used. The rotor selection can only be changed when the centrifuge is at a standstill.
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 6 Using the centrifuge → 6.3.2.9 Acceleration and deceleration curves Acceleration This function is used to select an acceleration curve. The system offers 10 fixed, programmed acceleration curves (curves 0-9).
WARNING The cycle display will be reset after the rotor and buckets have been replaced by the service department of Sigma Laborzentrifugen GmbH (see chapter 7.3 - "Service contact"). Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb...
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 6 Using the centrifuge 6.3.2.12 Input lock In order to prevent the centrifuge from being manipulated by unauthorised persons, inputs can be disabled via the menu. Inputs via the keypad, i.e. for starting or stopping a centrifugation run or for opening the lid, are enabled.
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 6 Using the centrifuge → → 6.3.3.1 Saving a program This function is only available if the centrifuge is at a standstill. NOTE • Enter the parameters that are to be included in the program.
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 6 Using the centrifuge Use as a robot-controlled centrifuge In addition to manual operation, the centrifuge is also suitable for being loaded by a robot. A separate computer is used for the control.
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 6 Using the centrifuge 6.4.3 Input The centrifuge can be controlled with several input commands. The input commands are divided into several categories: • Check commands, these commands trigger an immediate action •...
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 6 Using the centrifuge → Switching the centrifuge off • Open the centrifuge when it is not in use so moisture can evaporate. • Switch the centrifuge off by pressing the mains power switch.
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 7 Malfunctions and error correction → Malfunctions and error correction → General malfunctions Error messages are displayed as "Error" followed by a code number. If the acoustic signal is activated, it sounds when the error message is displayed.
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 7 Malfunctions and error correction 7.1.1 Emergency lid release In the event of a power failure, it is possible to manually open the centrifuge lid.
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 7 Malfunctions and error correction → Table of error codes Error no. Kind of error Measures Note • System error Allow to slow down All these errors stop the centrifuge or cause it to •...
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 7 Malfunctions and error correction → Service contact In the event of queries, malfunctions, or spare part enquiries: From Germany: Contact Sigma Laborzentrifugen GmbH An der Unteren Söse 50 37520 Osterode (Germany) Tel.
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 8 Maintenance and service → Maintenance and service The centrifuge, rotor, and accessories are subject to high mechanical stress. Thorough maintenance performed by the user extends the service life and prevents premature failure.
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 8 Maintenance and service → 8.1.1.1 Condenser (only refrigerated centrifuges with an air-cooled refrigeration system) In order to cool the refrigerant that is compressed by the refrigeration unit, centrifuges with an air-cooled refrigeration system use a lamellar condenser.
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 8 Maintenance and service 8.1.2.1 Plastic accessories The chemical resistance of plastic decreases with rising temperatures (see chapter 11.5 - "Resistance data"). • If solvents, acids, or alkaline solutions have been used, clean the plastic accessories thoroughly.
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 8 Maintenance and service → 8.1.4 Load bearing bolts Only greased load-bearing bolts ensure a uniform swing-out of the buckets and, therefore, the smooth operation of the centrifuge. Load-bearing bolts that are insufficiently greased may cause the centrifuge to stop due to an imbalance.
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 8 Maintenance and service → 8.1.5 Glass breakage In the case of glass breakage, immediately remove all glass particles (e.g. with a vacuum cleaner). Replace the rubber cushions since even thorough cleaning will not remove all glass particles.
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 8 Maintenance and service → 8.2.1 Autoclaving The service life of the accessories essentially depends on the frequency of autoclaving and use. • Replace the accessories immediately when the parts show changes in colour or structure or in the occurrence of leaks etc.
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 8 Maintenance and service → Service In the event of service work that requires the removal of the panels, there is a risk of electric shock or mechanical injury.
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 8 Maintenance and service → Return of defective centrifuges or parts Although we exercise great care during the production of our products, it may be necessary to return a unit or accessory to the manufacturer.
Page 63
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 8 Maintenance and service The defective part/unit must be packaged in a transport-safe manner. Please use the original packaging for the unit, if at all possible.
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 9 Disposal → Disposal → Disposal of the centrifuge In accordance with the directive 2012/19/EU, SIGMA centrifuges are marked with the symbol shown to the left. This symbol means that it is not permissible to dispose of the unit among household waste.
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 10 Technical data → 10 Technical data Manufacturer Sigma Laborzentrifugen GmbH An der Unteren Söse 50 37520 Osterode (Germany) Type: 4-5KRL 4-5KRL IVD Part no.:...
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 10 Technical data → 10.1 Ambient conditions • The figures are valid for an ambient temperature of +23°C and a nominal voltage 10 %. The minimum temperature is ≤ +4°C and depends on the rotor type, speed, and ambient temperature.
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 11 Appendix → 11 Appendix 11.1 Suitable accessories For use in a robot-controlled centrifuge, special equipment is required. Please contact our sales department: SIGMA Laborzentrifugen GmbH An der Unteren Söse 50...
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 11 Appendix → 11.2 Speed-gravitational-field-diagram Fig. 34: Speed-gravitational-field-diagram Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb 68 / 113 Translation of the original operating manual,...
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 11 Appendix → 11.3 Acceleration and deceleration curves Linear curves are numbered in the direction of increasing acceleration (from right to left). The deceleration curves are inverted images of the acceleration curves and are assigned the same numbers.
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 11 Appendix → 11.4 Table of the service life of rotors and accessories • The rotors and accessories must be put out of service after 10 years. Any use after this period may be permissible in individual cases after an inspection performed by the manufacturer.
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 11 Appendix → 11.5 Resistance data The data refer to resistance at 20°C. NOTE - no data 1 resistant 2 practically resistant 3 partially resistant...
Page 72
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 11 Appendix - no data 1 resistant 2 practically resistant 3 partially resistant 4 not resistant Medium Formula Chromic acid Chromic potassium KCr(SO saturated sulphate...
Page 73
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 11 Appendix - no data 1 resistant 2 practically resistant 3 partially resistant 4 not resistant Medium Formula Isopropyl alcohol Lactic acid Magnesium chloride MgCl...
Page 74
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 11 Appendix - no data 1 resistant 2 practically resistant 3 partially resistant 4 not resistant Medium Formula Sodium bisulphite NaHSO Sodium carbonate Sodium chloride...
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 11 Appendix 11.6 Serial Control Interface Specification Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb 75 / 113 Translation of the original operating manual,...
Page 76
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 11 Appendix Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb 76 / 113 Translation of the original operating manual,...
Page 77
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 11 Appendix Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb 77 / 113 Translation of the original operating manual,...
Page 78
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 11 Appendix Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb 78 / 113 Translation of the original operating manual,...
Page 79
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 11 Appendix Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb 79 / 113 Translation of the original operating manual,...
Page 80
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 11 Appendix Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb 80 / 113 Translation of the original operating manual,...
Page 81
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 11 Appendix Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb 81 / 113 Translation of the original operating manual,...
Page 82
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 11 Appendix Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb 82 / 113 Translation of the original operating manual,...
Page 83
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 11 Appendix Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb 83 / 113 Translation of the original operating manual,...
Page 84
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 11 Appendix Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb 84 / 113 Translation of the original operating manual,...
Page 85
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 11 Appendix Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb 85 / 113 Translation of the original operating manual,...
Page 86
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 11 Appendix Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb 86 / 113 Translation of the original operating manual,...
Page 87
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 11 Appendix Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb 87 / 113 Translation of the original operating manual,...
Page 88
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 11 Appendix Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb 88 / 113 Translation of the original operating manual,...
Page 89
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 11 Appendix Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb 89 / 113 Translation of the original operating manual,...
Page 90
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 11 Appendix Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb 90 / 113 Translation of the original operating manual,...
Page 91
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 11 Appendix Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb 91 / 113 Translation of the original operating manual,...
Page 92
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 11 Appendix Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb 92 / 113 Translation of the original operating manual,...
Page 93
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 11 Appendix Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb 93 / 113 Translation of the original operating manual,...
Page 94
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 11 Appendix Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb 94 / 113 Translation of the original operating manual,...
Page 95
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 11 Appendix Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb 95 / 113 Translation of the original operating manual,...
Page 96
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 11 Appendix Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb 96 / 113 Translation of the original operating manual,...
Page 97
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 11 Appendix Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb 97 / 113 Translation of the original operating manual,...
Page 98
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 11 Appendix Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb 98 / 113 Translation of the original operating manual,...
Page 99
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 11 Appendix Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb 99 / 113 Translation of the original operating manual,...
Page 100
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 11 Appendix Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb 100 / 113 Translation of the original operating manual,...
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 11 Appendix Lars 11.7 EC declaration of conformity Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb 101 / 113 Translation of the original operating manual,...
Page 102
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 11 Appendix Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb 102 / 113 Translation of the original operating manual,...
Page 103
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 11 Appendix Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb 103 / 113 Translation of the original operating manual,...
Page 104
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 11 Appendix Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb 104 / 113 Translation of the original operating manual,...
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 11 Appendix Declaration of conformity – China RoHS 2 11.8 Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb 105 / 113 Translation of the original operating manual,...
Page 106
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 11 Appendix Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb 106 / 113 Translation of the original operating manual,...
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 11 Appendix 11.9 UKCA declaration of conformity Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb 107 / 113 Translation of the original operating manual,...
Page 108
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD 11 Appendix Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb 108 / 113 Translation of the original operating manual,...
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD Index 12 Index Changes in structure ........60 Check commands ........49 Acceleration ..........44 Chemical and biological safety ....24 Acceleration curves ......44, 69 Chemical resistance of plastic ....
Page 110
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD Index Dimensions ..........29 Highly corrosive substances ...... 23 Direct hazard to the life and health ..... 18 Directive 2002/96/EC ......... 64 Imbalance ..........34, 36 Disinfectants ..........
Page 111
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD Index Mains switch ..........11 Power supply ..........31 Mains voltage ..........22 Preselection of a curve....... 44 Maintenance ..........55 Pressure (refrigerant) ......... 65 Malfunctions and error correction ....
Page 112
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD Index Rotors, cleaning and care ......57 Stopping of the centrifuge due to an imbalance ..........58 RS232 cable ..........10 Storage and transport ........ 29 Runtime .............
Page 113
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD Index Wear ............61 Weight ............29 Warranty and liability ........9 Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb 113 / 113 Translation of the original operating manual,...