8 Maintenance and service
8.1.4

Load bearing bolts

58 / 113
Only greased load-bearing bolts ensure a uniform swing-out of the buckets
and, therefore, the smooth operation of the centrifuge. Load-bearing bolts
that are insufficiently greased may cause the centrifuge to stop due to an
imbalance.
Clean the load-bearing bolts and bucket groove in order to remove the
old grease.
Apply a small amount of heavy-duty grease for load-bearing bolts (ref.
no. 71401, see the following picture) to both load-bearing bolts of a
bucket.
Fig. 32: Sufficient quantity of grease for one bolt
Install the bucket and swing it manually back and forth once in order to
distribute the grease.
Repeat this process with all the other buckets.