Using The Centrifuge; Initial Start-Up; Switching The Centrifuge On; Opening And Closing The Lid - Sigma 4-5KRL Manual

Table of Contents

Advertisement

6

Using the centrifuge

6.1

Initial start-up

DANGER
6.2

Switching the centrifuge on

NOTE
6.2.1

Opening and closing the lid

WARNING
Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb
Translation of the original operating manual,
Before the initial start-up, please ensure that your centrifuge is
properly set up and installed (see chapter 5 - "Set-up and
connection").
Press the mains power switch.
The display then illuminates. The centrifuge is ready for operation.
The start key remains inactive and can only be activated via the
connected PC or by entering a code.
Entering the code
Press and hold the stop key while entering the code.
Turn the function knob 1 notch to the left and then 3 nothes to the right.
Release the stop key.
The start key will be illuminated, the centrifuge is ready for operation.
The lid can be opened if the centrifuge is at a standstill and if the lid key is
illuminated.
Press the lid key in order to open the lid.
The centrifuge cannot be started if the lid is opened.
To close, press with both hands slightly on the lid until the lock is
audibly locked.
Do not place your fingers between the lid and the housing when closing
the lid. Risk of crushing!