6.4.3

Input

NOTE
6.4.3.1

Positioning of the rotor

Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb
Translation of the original operating manual,
The centrifuge can be controlled with several input commands. The input
commands are divided into several categories:
Check commands, these commands trigger an immediate action
Change of input values
Display of current process values
Display of current input values
Indication of the centrifuge status
Display of parameters
A complete list of these commands and several examples can be found in
chapter 11.6 - "Serial Control Interface Specification"
If the centrifuge is loaded and unloaded by way of a robot, special rotors
are used (see chapter 6.4.2 - "Rotors"). These rotors are positioned via an
input command.
The command "setpos n" moves the rotor automatically to the desired
position where it is then locked. During the positioning process, the
message "Posi" will be displayed in the upper area of the speed/RCF field
while a dynamic progress bar below this area indicates the progress or
duration of the process. The command that is to be executed will be
displayed in the upper area of the time and temperature field while the
lower area shows the selected rotor position.
Fig. 29: Indication during the positioning process
When the positioning process has been completed and the rotor has been
locked in the selected position, the message "Lock" will be displayed in the
speed/RCF field. The rotor is now in the correct position.
Fig. 30: Indication after the completion of the positioning process
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD
6 Using the centrifuge
49 / 113