Resistance Of Plastics; Service Life (For Centrifuges According To Machinery Directive) - Sigma 4-5KRL Manual

Table of Contents

Advertisement

3.6.7

Resistance of plastics

NOTE
3.6.8

Service life (for centrifuges according to Machinery Directive)

WARNING
NOTE
Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb
Translation of the original operating manual,
Chemical influences have a strong effect on the polymeric chains of
plastics, and, therefore, on their physical properties. Plastic parts can be
damaged if solvents, acids, or alkaline solutions are used.
Refer to the resistance data (see chapter 11.5 - "Resistance data")!
The rotors and accessories have a limited service life.
Perform regular checks (at least once per month) for safety reasons!
Pay special attention to changes, such as corrosion, cracks, material
abrasion, etc.
After 10 years, they must be inspected by the manufacturer.
After 50,000 cycles, the rotor must be scrapped for reasons of safety.
If other data concerning the service life are engraved on the rotor or
bucket, these data shall apply accordingly. For example, a bucket with
the engraving "max. cycles = 10,000" has a service life of 10,000
cycles, and a rotor with the engraving "Exp. date 01/27" must be
scrapped in January 2027 at the latest (see figure).
If a specification concerning the maximum number of cycles and a
specification concerning the service life (i.e. a date) are provided, the
specification that occurs first shall apply.
Fig. 6: Different service life – engraving on the bucket/rotor
Refer to the table of the service life of rotors and accessories (see
chapter 11.4 - " Table of the service life of rotors and accessories ")!
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD
3 Safety
25 / 113

Advertisement

Table of Contents
loading

This manual is also suitable for:

4-5krl ivd9130991564100008100009

Table of Contents