The centrifuge is packaged in a wooden crate.
After taking off the lid, remove the side panels.
Remove the packaging material.
Lift the centrifuge upwards with a lifting device to lift it safely. When
lifting the centrifuge, always reach under the centrifuge from the side.
The centrifuge weighs approx. 135 kg!
Retain the packaging for any possible future transport of the centrifuge.
The transport safety device consists of two hexagon socket screws which
are located at the bottom panel.
The transport safety device must be removed prior to start-up because the
screws lock the motor bearings!
Removal
Lift the centrifuge upwards at the front side. Always reach under from
the side.
Put a suitable object, e.g. a wooden block, between the table and centri-
fuge. The two screws can now be seen at the bottom panel.
Fig. 8: Locking screws (transport safety device)
Unscrew the hexagon socket screws with a hexagon socket wrench
(size 4, included in the scope of supply) anti-clockwise.
Retain the transport safety device for the possibility of the return of the
centrifuge.