Refrigerated Centrifuge for Robot Use Sigma 4-5KRL
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD
11 Appendix
→
11.4
Table of the service life of rotors and accessories
•
The rotors and accessories must be put out of service after 10 years. Any use after this period
may be permissible in individual cases after an inspection performed by the manufacturer.
•
If a specification concerning the maximum number of cycles and a specification concerning the
service life (i.e. a date) are provided, the specification that occurs first shall apply.
•
After 50,000 cycles, rotors must be scrapped for safety reasons.
Rotor / bucket Cycles Service life
9100
9366
11805
11806
12082
12083
12084
12085
12092
12093
12094
12096
12097
12134
12135
12137
12500
12600
13218
13221
13296
13299
13635
13650
13845
13850
13860
91060
70 / 113
Autoclaving Suitable for centrifuge
("Exp.Date")
35,000
15,000
10 years
10 years
7 years
7 years
7 years
7 years
5 years
5 years
5 years
5 years
5 years
5 years
5 years
5 years
7 years
7 years
20,000
10,000
35,000
5 years
5 years
25,000
20,000
20,000
10,000
10 years
15,000
10 years
10,000
4-16S, 4-16KS, 4-16KHS,
6-16S, 6-16HS, 6-16KS,
6-16KHS
4-5KL, 4-16S, 4-16KS,
4-16KHS, 6-16S, 6-16HS,
6-16KS, 6-16KHS
8KS, 8KBS
8KS, 8KBS
1-14, 1-14K
1-14, 1-14K
1-14, 1-14K
1-14, 1-14K
20x
1-14, 1-14K
20x
1-14, 1-14K
20x
1-14, 1-14K
20x
1-14, 1-14K
20x
1-14, 1-14K
20x
1-16, 1-16K
20x
1-16, 1-16K
20x
1-16, 1-16K
6-16S, 6-16HS,
6-16KS, 6-16KHS
6-16S, 6-16HS,
6-16KS, 6-16KHS
4-16S, 4-16KS, 4-16KHS,
6-16S, 6-16HS,
6-16KS, 6-16KHS
4-16S, 4-16KS, 4-16KHS,
6-16S, 6-16HS,
6-16KS, 6-16KHS
2-7, 2-16P, 2-16KL,
10x
2-16KHL
2-7, 2-16P, 2-16KL,
10x
2-16KHL, 3-30KS,
3-30KHS
6-16S, 6-16HS,
6-16KS, 6-16KS
6-16S, 6-16HS,
6-16KS, 6-16KHS
8KS
8KS
8KBS
6-16S, 6-16HS
Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb
Translation of the original operating manual,
Remarks
Special software
required