6 Using the centrifuge

6.3.2.12 Input lock

NOTE
6.3.3

Program mode

46 / 113
In order to prevent the centrifuge from being manipulated by unauthorised
persons, inputs can be disabled via the menu. Inputs via the keypad, i.e. for
starting or stopping a centrifugation run or for opening the lid, are enabled.
Activating a simple input lock:
Position the cursor over the symbol "
of the display.
As long as the symbol is displayed, the parameters of the centrifuge cannot
be changed.
Activating a permanent input lock:
Press the start key three times and hold for approximately 2 seconds
when pressing it for the third time.
After the activation of this function, the padlock symbol flashes. The input
lock is activated.
Proceed in the same manner in order to deactivate the permanent input
lock.
Fig. 25: "Padlock" symbol indicating an activated input lock
If the centrifuge is robot-controlled via the computer, the display can only
be used for display purposes. Only the stop key is active at all times and
can be used to switch the centrifuge off in an emergency situation.
A program contains all the data that are required for a centrifuge run.
Certain sedimentation results can be repeated under identical conditions.
Programs can be saved, loaded, executed, and edited when the centrifuge
is at a standstill.
A maximum of 50 programs can be stored under the numbers 1 - 50. The
rapid cooling program "Rapid Temp" does not occupy any storage location
and cannot be deleted. It is used to cool the centrifuge without any vessels.
"--" means that the values that are currently set are not a stored program.
The programs can be protected against unauthorised use, modification, or
deletion with the aid of an input lock (see chapter 6.3.2.12 - "Input lock").
" in the lower right-hand corner
Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb