8 Maintenance and service

8.2.1 Autoclaving

NOTE
Accessories
Aluminium buckets
Aluminium rotors
Glass tubes
Polyallomer / polycarbonate rectangular carriers
Polyallomer / polycarbonate round carriers
Polyamide buckets
Polycarbonate / polyallomer lids for angle rotors
Polycarbonate tubes
Polyphenylsulfone (PPSU) caps for buckets
Polypropylene balance adapter for blood-bag systems
Polypropylene copolymer tubes
Polypropylene rectangular carriers
Polypropylene rotors
Polypropylene round carriers
Polysulfone caps for buckets
Polysulfone lids for angle rotors
Rubber adapters
Stainless-steel balance weight for blood-bag systems
Teflon tubes
60 / 113
The service life of the accessories essentially depends on the frequency of
autoclaving and use.
Replace the accessories immediately when the parts show changes in
colour or structure or in the occurrence of leaks etc.
During autoclaving, the caps of the tubes must not be screwed on in
order to avoid the deformation of the tubes.
It cannot be excluded that plastic parts, e.g. lids or carriers, may deform
during autoclaving.
Max. temp.
Min. time
(°C)
(min)
134-138
3
134-138
3
134-138
3
115-118
30
115-118
30
115-118
30
115-118
30
115-118
30
134-138
3
115-118
30
115-121
30
115-118
30
115-118
30
115-118
30
134-138
3
134-138
3
115-118
30
121
30
134-138
3
Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb
Translation of the original operating manual,
Max. time
Max. cycles
(min)
5
-
5
-
40
-
40
-
40
-
40
10
40
20
40
20
5
100
40
n.s.
40
20
40
-
40
20
40
-
5
100
5
100
40
-
30
n.s.
5
100