Refrigerated Centrifuge for Robot Use Sigma 4-5KRL
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD
6 Using the centrifuge
6.2.2.3
Adapters
→
6.2.2.4
Vessels
36 / 113
In order to ensure easy handling, even if vessels of various sizes are used,
carrier systems were developed.
•
Load the opposite adapters with the same number of vessels and with
the same weights in order to avoid imbalance.
•
If all of the compartments of a carrier are not used, the buckets must be
loaded evenly. Loading the edges of a bucket only is not permissible.
•
Load the vessels outside of the centrifuge. Liquids in the buckets or
multiple carriers cause corrosion.
•
Fill the vessels carefully and arrange them according to their weight.
Imbalances result in the excessive wear of the bearings.
•
Always fill the tubes up to their useful volume (= the volume that is
stated for the tube).
•
When using glass vessels, the maximum value of 4,000 x g must not be
exceeded (except special high-strength glass vessels; please refer to
the information provided by the manufacturer).
•
After the centrifugation, remove the vessels carefully in order to prevent
the samples from mixing.
•
Follow the safety instructions and hazard warnings (see chapter 3 -
"Safety")!
Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb
Translation of the original operating manual,