6 Using the centrifuge
6.2.2.3

Adapters

6.2.2.4

Vessels

36 / 113
In order to ensure easy handling, even if vessels of various sizes are used,
carrier systems were developed.
Load the opposite adapters with the same number of vessels and with
the same weights in order to avoid imbalance.
If all of the compartments of a carrier are not used, the buckets must be
loaded evenly. Loading the edges of a bucket only is not permissible.
Load the vessels outside of the centrifuge. Liquids in the buckets or
multiple carriers cause corrosion.
Fill the vessels carefully and arrange them according to their weight.
Imbalances result in the excessive wear of the bearings.
Always fill the tubes up to their useful volume (= the volume that is
stated for the tube).
When using glass vessels, the maximum value of 4,000 x g must not be
exceeded (except special high-strength glass vessels; please refer to
the information provided by the manufacturer).
After the centrifugation, remove the vessels carefully in order to prevent
the samples from mixing.
Follow the safety instructions and hazard warnings (see chapter 3 -
"Safety")!
Version 11/2017, Rev. 1.7.1 of 08/12/2022 • sb