Self-Tests 10xx: A7 Reference (Option UNJ) Self-Test Errors

Figure 1-8
Before proceeding to the reported self-test error code, check the voltages in Table1-25. If any
voltages are out of specification troubleshoot the supply problem first.
Table 1-25
Supply
Voltage
(Vdc)
1000 Power Supply
1. Replace the A7 Reference.
1-36
Connector
Pins
+15
P42-6, 21
-15
P42-2, 17
+9
P42-5, 20
Minimum
Maximum
Value
Value
(Vdc)
(Vdc)
+14.55
+15.45
-14.55
-15.45
+8.82
+9.18
Origin
Main Supply
Main Supply
YIG Driver