A6 Frac-N; Tools Required; Removal Procedure; Replacement Procedure - Agilent Technologies PSG Series Service Manual

Signal generator
Hide thumbs Also See for PSG Series:
Table of Contents

Advertisement

Assembly Replacement

A6 Frac-N

A6 Frac-N

Tools Required

none

Removal Procedure

Refer to Figure 2-10 for this procedure.
1. Disconnect the power cord.
2. Remove the covers from the signal generator. Refer to
"Inner Instrument Cover" on page
3. Position the signal generator with the front panel facing you.
4. Lift the retention levers and disconnect the A6 Frac-N from the A31 Motherboard.
5. Lift the A6 Frac-N out of the slot.

Replacement Procedure

Reverse the order of the removal procedure.
Perform the post-repair adjustments and performance tests that pertain to this removal procedure.
2-22
"Outer Instrument Cover" on page 2-4
2-6.
and

Advertisement

Table of Contents
loading

This manual is also suitable for:

E8241aE8244aE8251aE8254a

Table of Contents