Self-Tests 7xx: A6 Frac-N Self-Test Errors

Figure 1-6
Before proceeding to the reported self-test error code, check the voltages in Table1-22. If any
voltages are out of specification troubleshoot the supply problem first.
Table 1-22
Supply
Voltage
(Vdc)
700 Ref Voltage
1. Replace the A6 Frac-N.
Connector
Pins
+32
P32-7
+15
P32-6, 21
-15
P32-2, 17
+5.2
P32-14, 29
+9
P32-5, 20
-5.2
P32-4,19
-6
P32-3, 18
Troubleshooting Assembly Level Problems
Minimum
Maximum
Value
Value
(Vdc)
(Vdc)
+31.04
+32.96
+14.55
+15.45
-14.55
-15.45
+5.04
+5.37
+8.82
+9.18
-5.1
-5.3
-5.88
-6.12
Troubleshooting
Origin
Main Supply
Main Supply
Main Supply
Main Supply
YIG Driver
YIG Driver
YIG Driver
1-31