Assembly Replacement

A11 Pulse/Analog Modulation Generator

A11 Pulse/Analog Modulation Generator

Tools Required

needle-nose pliers

Removal Procedure

Refer to Figure 2-15 for this procedure.
1. Disconnect the power cord.
2. Remove the covers from the signal generator. Refer to
"Inner Instrument Cover" on page
3. Position the signal generator with the front panel facing you.
4. Using the needle-nose pliers, disconnect the W1 flexible cable from A11J401.
5. Using the needle-nose pliers, disconnect the W2 flexible cable from A11J402.
6. Using the needle-nose pliers, disconnect the W3 flexible cable from A11J403.
7. Using the needle-nose pliers, disconnect the W5 flexible cable from A11J201.
8. Using the needle-nose pliers, disconnect the W6 flexible cable from A11J203.
9. Using the needle-nose pliers, disconnect the W7 flexible cable from A11J202.
10. Using the needle-nose pliers, disconnect the W16 flexible cable from A11J204.
11. Using the needle-nose pliers, disconnect the W17 flexible cable from A11J205.
12. Lift the retention levers and disconnect the A11 Pulse/Analog Modulation Generator from the
A31 Motherboard.
13. Lift the A11 Pulse/Analog Modulation Generator out of the slot.

Replacement Procedure

Reverse the order of the removal procedure.
Perform the post-repair adjustments and performance tests that pertain to this removal procedure.
2-32
"Outer Instrument Cover" on page 2-4