Self-Tests 9Xx: A10 Alc Self-Test Errors - Agilent Technologies PSG Series Service Manual

Signal generator
Hide thumbs Also See for PSG Series:
Table of Contents

Advertisement

Troubleshooting
Troubleshooting Assembly Level Problems

Self-Tests 9xx: A10 ALC Self-Test Errors

Figure 1-7
Before proceeding to the reported self-test error code, check the voltages in Table1-24. If any
voltages are out of specification troubleshoot the supply problem first.
Table 1-24
Supply
Voltage
(Vdc)
+5.2 Digital high
+3.4 Digital Low
1-34
Connector
Pins
+32
P122-4
+15
P122-28
-15
P122-3,
+5.2
P122-2, 27
-5.2
P122-1, 26
P121-64, 65,
129, 130
P121-60, 61,
62, 63, 125,
126, 127, 128
Minimum
Maximum
Value
Value
(Vdc)
(Vdc)
+31.04
+32.96
+14.55
+15.45
-14.55
-15.45
+5.04
+5.37
+5.04
+5.37
+5.04
+5.36
+3.29
+3.5
Origin
Main Supply
Main Supply
Main Supply
Main Supply
Main Supply
Main Supply
Main Supply

Advertisement