Assembly Replacement

A26 MID (Microcircuit Interface Deck)

A26 MID (Microcircuit Interface Deck)

Tools Required

T-10 driver

Removal Procedure

Refer to Figure 2-29 for this procedure.
1. Disconnect the power cord.
2. Remove the covers from the signal generator. Refer to
"Inner Instrument Cover" on page
3. Position the signal generator with the A31 Motherboard facing you and the RF deck on top.
4. Disconnect the W28 ribbon cable from A26J15.
5. If you have a 40 GHz signal generator, disconnect the W29 ribbon cable from A23J33.
6. Disconnect the W20 ribbon cable from A26J20.
7. Disconnect the W33 ribbon cable from A26J32.
8. Disconnect the W34 ribbon cable from A26J31.
9. Disconnect the W35 ribbon cable from A26J10.
10. Disconnect the W36 ribbon cable from A26J13.
11. Disconnect the A25BW1 power cable from A26J12.
12. Disconnect the W19 ribbon cable from A26J30.
13. Using the T-10 driver, loosen the screw that secures the A26 MID to the RF deck.
14. Slide the A26 MID towards the front panel until the locking standoffs are free.
15. Remove the A26 MID from the RF bracket.

Replacement Procedure

Reverse the order of the removal procedure.
Torque all T-10 screws to 9 in-lbs.
Perform the post-repair adjustments and performance tests that pertain to this removal procedure.
2-60
"Outer Instrument Cover" on page 2-4
2-6.
and