A18Bt1; Tools Required; Removal Procedure; Replacement Procedure - Agilent Technologies PSG Series Service Manual

Signal generator
Hide thumbs Also See for PSG Series:
Table of Contents

Advertisement

Assembly Replacement

A18BT1

A18BT1

WARNING
This battery contains lithium. Do not incinerate or puncture this battery. Do not install this battery
backwards. To dispose of the battery in a safe manner, refer to
Disposal," on page

Tools Required

flat head screw driver

Removal Procedure

Refer to Figure 2-16 for this procedure.
1. Disconnect the power cord.
2. Remove the covers from the signal generator. Refer to
"Inner Instrument Cover" on page
3. Position the signal generator with the front panel facing you.
4. Disconnect the W18 ribbon cable A18J1.
5. Disconnect the W19 ribbon cable from A18J2.
6. Lift the retention levers and disconnect the A18 CPU from the A31 Motherboard.
7. Lift the A18 CPU out of the slot.
8. Using the flat head screw driver, remove the A18BT1 by leveraging the battery out of the battery's socket.

Replacement Procedure

To install the battery, the positive side is aligned with the positive sign on the A18's battery clip.
Reverse the remaining steps of the removal procedure.
Perform the post-repair adjustments and performance tests that pertain to this removal procedure.
2-36
5-4.
"Outer Instrument Cover" on page 2-4
2-6.
Chapter 5, "Lithium Battery
and

Advertisement