1001 10 MHz Test
1. Using an oscilloscope, check P41-2 for a 10 MHz signal greater than 1.12 V
replace the
A7 Reference.
2. If the signal is not present, replace the internal 10 MHz standard.
1002 1 GHz Test
1. Replace the A7 Reference.
Troubleshooting Assembly Level Problems
. If the signal is present,
p-p
Troubleshooting
1-37