A18 Cpu; Tools Required; Removal Procedure; Replacement Procedure - Agilent Technologies PSG Series Service Manual

Signal generator
Hide thumbs Also See for PSG Series:
Table of Contents

Advertisement

Assembly Replacement

A18 CPU

A18 CPU

Tools Required

none

Removal Procedure

Refer to Figure 2-16 for this procedure.
1. Disconnect the power cord.
2. Remove the covers from the signal generator. Refer to
"Inner Instrument Cover" on page
3. Position the signal generator with the front panel facing you.
4. Disconnect the W18 ribbon cable from A18J1.
5. Disconnect the W19 ribbon cable from A18J2.
6. Lift the retention levers and disconnect the A18 CPU from the A31 Motherboard.
7. Lift the A18 CPU out of the slot.

Replacement Procedure

Reverse the order of the removal procedure.
Perform the post-repair adjustments and performance tests that pertain to this removal procedure.
2-34
"Outer Instrument Cover" on page 2-4
2-6.
and