A9 Yig Driver; Tools Required; Removal Procedure; Replacement Procedure - Agilent Technologies PSG Series Service Manual

Signal generator
Hide thumbs Also See for PSG Series:
Table of Contents

Advertisement

Assembly Replacement

A9 YIG Driver

A9 YIG Driver

Tools Required

needle-nose pliers

Removal Procedure

Refer to Figure 2-13 for this procedure.
1. Disconnect the power cord.
2. Remove the covers from the signal generator. Refer to
"Inner Instrument Cover" on page
3. Position the signal generator with the front panel facing you.
4. Using the needle-nose pliers, disconnect the W10 flexible cable from the A9J3.
5. Using the needle-nose pliers, disconnect the W11 A9J4 flexible cable from the A9J4.
6. Disconnect the W12 ribbon cable from the A9J5.
7. Lift the retention levers and disconnect the A9 YIG Driver from the A31 Motherboard.
8. Lift the A9 YIG Driver out of the slot.

Replacement Procedure

Reverse the order of the removal procedure.
Perform the post-repair adjustments and performance tests that pertain to this removal procedure.