A3 Power Switch; Tools Required; Removal Procedure; Replacement Procedure - Agilent Technologies PSG Series Service Manual

Signal generator
Hide thumbs Also See for PSG Series:
Table of Contents

Advertisement

Assembly Replacement

A3 Power Switch

A3 Power Switch

Tools Required

T-10 driver

Removal Procedure

Refer to Figure 2-7 for this procedure.
1. Disconnect the power cord.
2. Remove the front panel. Refer to
3. Disconnect the W8 ribbon cable from the A3 Power Switch by lifting the white tabs on the connector.
4. Using the T-10 driver, remove the four screws (12) from the A3 Power Switch.
5. Remove the A3 Power Switch.

Replacement Procedure

Reverse the order of the removal procedure.
Torque all T-10 screws to 9 in-lbs.
Perform the post-repair adjustments and performance tests that pertain to this removal procedure.
2-16
"Front Panel" on page
2-8.