•press Frequency - set signal generator to frequency in problem area
•press Amplitude - set to +25 dBm
•press RF ON/OFF - turn RF On
•turn ALC Off
•Power Search Manual
•Do Power Search
4. Using the RF Path Block Diagram, check the power levels at the points shown.
5. Once the problem is located, verify any cable between assemblies is good before replacing the assembly.
6. To measure the A6 Frac-N output level, with the signal generator on, remove the A8 Output and probe the
right hand mmx connector's center pin. The power level should be ≥+5 dBm.
WARNING
Always turn the signal generator off before installing any assembly.
7. If power levels look good through the path in ALC Off mode, the problem is most likely in the ALC loop. Go
to troubleshooting the ALC loop.

Troubleshooting Pulse Modulation

1. Run a full self-test and troubleshoot any reported failures before proceeding.
2. Determine the frequency of the pulse modulation problem.
3. Configure the signal generator to the pulse modulation setting that produces the problem.
4. Connect an oscilloscope to the LF OUTPUT connector. The waveform should be present on the oscilloscope.
5. If the waveform is present, continue to step 7.
6. If the waveform is not present or the signal is not the correct amplitude or type, replace the A11
Pulse/Analog Modulation Generator.
7. Turn ALC Off. If in ALC On mode, the ALC bandwidth causes a reduction is pulse amplitude at pulse
widths <1µs. If turning ALC Off corrects the problem, then verify the user has set up pulse modulating in
either ALC Off or Power Search Modes.
Troubleshooting Problems <3.2 GHz
8. If the pulse modulation problem occurs at <3.2 GHz, check P52-23 on the A31 Motherboard. Compare the
results with Table1-37.
Table 1-37
+5 Vdc
9. If signal levels are good, replace the A8 Output.
10. If the signals are bad, replace the A11 Pulse/Analog Modulation Generator.
Pulse On
Pulse Off
0 to +5 Vdc (pulses at modulation rate)
Troubleshooting
Troubleshooting the RF Path
1-59