Assembly Replacement

A10 ALC

A10 ALC

Tools Required

needle-nose pliers

Removal Procedure

Refer to Figure 2-14 for this procedure.
1. Disconnect the power cord.
2. Remove the covers from the signal generator. Refer to
"Inner Instrument Cover" on page
3. Position the signal generator with the front panel facing you.
4. Using the needle-nose pliers, disconnect the W4 flexible cable from A10J4.
5. Using the needle-nose pliers, disconnect the W13 flexible cable from A10J201.
6. Using the needle-nose pliers, disconnect the W14 flexible cable from A10J5.
7. Using the needle-nose pliers, disconnect the W15 flexible cable from A10J3.
8. Lift the retention levers and disconnect the A10 ALC from the A31 Motherboard.
9. Lift the A10 ALC out of the slot.

Replacement Procedure

Reverse the order of the removal procedure.
Perform the post-repair adjustments and performance tests that pertain to this removal procedure.