Chapter 3
Before You Begin...
2. Inspect each item for damage.
3. Place the FOCUS GC on your bench, allowing space to the rear of the
WARNING! The FOCUS GC weighs 30 kg (70 lbs). Two people should lift the instrument onto the
benchtop.
You should already have configured your laboratory according to the space
requirements in Chapter 1,
supplies should be accessible.
Remove all protecting bags (if any) from the detector cell.
Optional equipment should be placed near the GC so you can connect it easily.
After placing the FOCUS GC on the bench, open the oven door and remove any
packing material or other debris.
Before continuing with the installation, you should determine the following:
Refer to
for a list of detector and injector gas requirements.
48
If equipment is damaged, keep boxes and their equipment in their existing
condition and immediately notify the carrier.
Submit a damage claim directly to the carrier, and send a copy (including
any shortage claims) to your authorized Thermo Fisher Scientific sales
representative.
Do not return any equipment to the dealer or the factory without prior
factory authorization.
instrument for venting.
the type of detector installed on the GC
the type of injector port installed on the GC
the type of carrier gas required for the detector
the type of makeup gas required for the detector
Gas and Plumbing Requirements
Laboratory
Preparation, and the gas and power
in Chapter 1,
Unpacking the Instrument
Laboratory
Preparation,
Site Preparation and Installation Manual