Power Requirements
Most UPS systems are not designed to provide high-quality distortion-free power for
scientific equipment.
The 120 V ac systems are fitted with U.S. standard National Electronics
Manufacturers Association (NEMA) 5-20P power plugs. A NEMA 5-20P power
plug and its corresponding outlet are rated at 20 A and 125 V ac.
The power cables from the personal computer, monitor, and printer are
approximately 2 m (6 ft) long. The 120 V ac systems are fitted with NEMA 5-15P
plugs. For optional instruments, the plug requirements may vary. Refer to your
product's user manual for specifications.
Figure 1-1 shows the NEMA power plugs and outlets.
The 230 V ac systems are fitted with Shuco German-type power plugs. Plug
requirements are dictated by country. Figure 1-2 shows the Shuco power plug and
outlet.