How to Attach a Swagelok Tee or Cross

To use a single gas source for more than one inlet or detector module, use a
Swagelok tee or cross to split the gas flow. Use the following procedure to
connect a Swagelok tee or cross:
1. Use a tubing cutter to cut gas supply tubing where you want to install the tee
2. Connect the tubing to the tee or cross with a Swagelok fitting, as described in
3. Measure the distance from the tee or cross to the inlets or detectors, and cut
4. Connect the tubing to the tee or cross ends with Swagelok fittings.
5. Install Swagelok caps on any open ends of the tee or cross that you do not
52
or cross.
How to Use Swagelok Tube Fittings
tubing in the appropriate lengths.
plan to connect with tubing.
on page 50.
Site Preparation and Installation Manual
Gas Plumbing Basics