Chapter 1
Laboratory Preparation
CAUTION
You should never connect the FOCUS GC and its peripheral devices to the same
electrical wall outlet. You will run the risk of overloading the circuit.
Tables 1-3 and 1-4 show maximum current requirements for the FOCUS GC
system and optional instruments.
NOTE
Other peripheral devices, such as purge and trap units, may be connected to the FOCUS
GC. These peripheral devices may have additional power requirements. Consult your local
Thermo Fisher Scientific CSE for assistance.
26

Table 1-3. Maximum Current Requirements for the FOCUS GC Standard System

Instrument
FOCUS GC
computer
monitor
Total

Table 1-4. Maximum Current Requirements for FOCUS GC Optional Instruments

Instrument
GC Autosamplers
AI/AS 3000
TriPlus
Printers
ink jet printer/plotter
laser printer
120 V Current
Requirement
(in amperes)
20
4
2
26
120 V Current
Requirement
(in amperes)
0.6
2
1.8
7
Site Preparation and Installation Manual
Power Requirements
230 V Current
Requirement
(in amperes)
10
2
1
13
230 V Current
Requirement
(in amperes)
0.3
1
0.9
3.5