6 .2 Potential errors

Independent of the priority group, all errors found are indicated with their relevant number.
In order to facilitate operation, the last error is stored in a non-volatile memory. The stored error may
be retrieved and deleted in the main menu.
If a combination of errors occurs simultaneously, they are indicated or stored in the error memory
observing the following sequence.
Priority group I
Error
Cause
No. 1
No system parameter available
No. 2
Incorrect checksum of parameter
memory
No. 3
Incorrect checksum of program
memory
No. 4
Incorrect checksum of data memory
No. 5
Internal controller error
Priority group II
Error
Cause
No. 10
Sensor not connected; cable
between FC100-CA and sensor
or sensor defective
Sensor selected (CONFIGURATION
menu) differs from sensor connected
No. 21
Medium temperature too high
60
ERRORS
Rectification
Return to supplier.
Return to supplier.
Return to supplier.
Return to supplier.
Return to supplier.
Rectification
Check cable or replace sensor.
Correct sensor selection in
CONFIGURATION menu