Potential Errors - Flow vision FC 01- CA User Manual

Table of Contents

Advertisement

8 .2 Potential errors

Independent of the priority group, all errors found are indicated with their relevant number.
In order to facilitate operation, the last error is stored in a non-volatile memory. The stored error
may be retrieved and deleted in the main menu.
If a combination of errors occurs simultaneously, they are indicated or stored in the error memory
observing the following sequence.
Priority group I
Error
Cause
No. 1
No system parameter available
No. 2
Incorrect test sum of parameter
memory
No. 3
Incorrect test sum of program
memory
No. 4
Incorrect test sum of data memory
No. 5
Internal controller error
Priority group II
Error
Cause
No. 10
Sensor not connected; or cable
between FC01-CA and sensor
defective; or defective sensor
Sensor selected (configuration)
differs from sensor connected
No. 21
Medium temperature too high
FC01- CA | Flow Meter
ERRORS
Rectification
Return to supplier.
Return to supplier.
Return to supplier.
Return to supplier.
Return to supplier.
Rectification
Check cable or replace sensor.
Correct sensor selection in
configuration menu
59

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the FC 01- CA and is the answer not in the manual?

Questions and answers

Table of Contents