Software Configuration
5.
When the alignment of the display is adjusted, retry the visual check task and verify that the
linearity has improved (See section "Run an individual task" for more information about
executing a single task). You can also check this without executing the task. Choose "Test
Patterns..." in the main view. A window with available test patterns shows up. Select "GEMS
Linearity-Geometry" from this list and click "Load". The linearity pattern is displayed on the
selected display of your configuration. Right-click on the test pattern for a popup menu with
further actions for this pattern.
5-6 Visual check – focus problem
1.
The focus problem cannot be solved with the current version of MediCal Pro. It is
recommended that you consult a service technician with a version of MediCal Maintenance to
adjust the focus of your display.
2.
When the focus of the display is adjusted, retry the visual check task and verify that the focus
has improved (See section "Run an individual task" for more information about executing a
single task). You can also check this without executing the task. Choose "Test Patterns..." in
the main view. A window with available test patterns shows up. Select "GEMS Focus" from this
list and click "Load". The focus pattern is displayed on the selected display of your
CNF 006 Monitor Calibration with Barco Tool (MediCal Pro)
Page 133