Software Configuration
22. Notice the graphic representation in the lower right-hand corner that shows the screen sensor
used in the test. Close MediCal Pro now.
Note:
- For 2 megapixel grayscale displays, the white is set to 65 fL, the black to 0.2 fL (DICOM as
display function with Ambient Light Correction).
- For 4 or 5 megapixel grayscale displays, white is set to 85 fL, black to 0.2 fL (DICOM as
display function with Ambient Light Correction).
- You can check these settings by right-clicking on a display and selecting Presets; there you
will see the target definitions.
- For non-Barco color displays, enter VALID data for the display brand name, type and serial
number. The white and black is measured on the next calibration (same display function
DICOM with Ambient Light Correction).
Section 4: Individual Maintenance Tasks with MediCal Pro
4-1 Q/A Check
1.
To execute all due tasks for today, go to Start > Programs > MediCal Pro > MediCal Pro.
2.01.
2.
Select Q/A Check in the main window.
3.
A message appears; select Yes to execute all due tasks for today and follow the instructions for
each individual task. Select No to skip the execution of the tasks. If there are no tasks to be
executed today, you are informed of this.
Note:
You can see the total result in the main window or look up individual results by selecting Q/A
Setup.