GE Centricity RA600 Service Manual page 131

Table of Contents

Advertisement

Software Configuration
2.
You need to be logged in as an advanced user. If you are not yet in advanced mode (your
configuration is not visible at the bottom of the main window), you need to log in again. Go to
Tools > Options > User Login... in the menu. The login window appears.
3.
Leave the user name as it is and type "advanced" as user password. Press OK. The advanced
view of MediCal Pro appears with the configuration of your displays.
4.
To start the alignment tool for a display, right-click on the picture of the display. Select
"Alignment..." in the popup menu.
Note:
Make sure the display where you do an alignment is the right one! Compare the serial number
of the display found in the task history with the serial number of the display you are about to
align. This serial number can be retrieved via the "Properties" option.
CNF 006 Monitor Calibration with Barco Tool (MediCal Pro)
Page 131