Software Configuration
5.
When the display is calibrated again, retry the visual check task and verify that the squares of
the SMPTE test pattern are visible now (See section "Run an individual task" for more
information about executing a single task). You can also check this without executing the task
by choosing "Test Patterns" in the main view. A window with available test patterns shows up.
Select "SMPTE" from this list and click "Load". The SMPTE pattern is displayed on the
selected display of your configuration. Right-click on the test pattern for a popup
5-5 Visual check – linearity problem
1.
In case the linearity of the display is problematic, it is recommended that you adjust your
display with the use of the alignment tool. Follow the procedure described below.
CNF 006 Monitor Calibration with Barco Tool (MediCal Pro)
Page 130