Software Configuration
3.
You need to be logged in as an advanced user. If you are not yet in advanced mode (your
configuration is not visible at the bottom of the main window), you need to log in again. Go to
Tools > Options > User Login ... in the menu. The login window appears.
4.
Leave the user name as it is and type "advanced" as user password. Press OK. The advanced
view of MediCal Pro appears with the configuration of your displays.
5.
To check the setting of the LUC of a display, right click on the picture of a display. Make sure
the option "Luminance Uniformity Correction..." is selected (The box in front of the option is
checked).
If the LUC was already enabled, do a Full Screen Calibration on this display. Select "Full
Screen Calibration" in the same popup menu and follow the instructions of the calibration.
If the LUC was disabled, enable it now. Calibration is not necessary.
Note:
Make sure the display you operate on is the correct one! Compare the serial number of the
display found in the task history with the serial number of the display you are manipulating in
the configuration view. This serial number can be retrieved via the "Properties..." option in the
popup menu of the display.
CNF 006 Monitor Calibration with Barco Tool (MediCal Pro)
Page 125