Software Configuration
6.
A message box may appear, asking you for a connection to MediCal Administrator. This is a
separate Barco network administration program for monitor calibration. Select No if you do not
use this program. An additional message then appears, describing where to find this
connection later; click OK.
7.
The following initial conifiguration window appears (example shown is for a "dual head" card
which controls one or two monitors). Click on Configuration Setup... to start the configuration
wizard..
8.
Click Next twice to start the wizard searching, first for Barco monitors, i.e. Barco display
controller card(s). The wizard reports its findings: If a Barco card is installed, the number of
displays connected is reported and wizard asks if this is correct. If it is correct, click Yes.
CNF 006 Monitor Calibration with Barco Tool (MediCal Pro)