5. User interface > 5.9. Controlling and monitoring the sonication

5.9.2. Sonication buttons

The Start Sonication and Abort Sonication buttons are used for starting and aborting sonication.
The buttons are active only in those stages where sonication is possible, i.e. during test sonication
and therapy.
The Abort Sonication button should be used to stop the treatment only in situations in which there
is no risk to the patient. In all other cases, use the stop button on the Safety Device.
A warning symbol is displayed on top of the Start Sonication button when the user is allowed to
override the remaining part of the system-recommended cooling time (see 4.5.4. Cooling time). An
orange warning indicates a very high-risk sonication Generally, it is not recommended to start the
next sonication until the entire system-recommended cooling time has passed.
Instructions for Use
109745C2 / 02-2022
Start Sonication
Abort Sonication
Override allowed
High-risk sonication, override not recommended