Storing The Qa Phantom - Profound Sonalleve MR-HIFU Instructions For Use Manual

Table of Contents

Advertisement

7. Maintenance > 7.2. Sonalleve MR-HIFU Quality Assurance (QA) procedure
The QA Test fails with the message
"Problem with QA Phantom"
The QA Test fails due to: "Heating
location check failed", "Maximum
temperature check failed", "Heating shape
check failed", "Heat increase check
failed", some other reason, or an
unknown reason
If repeating the QA Test does not help, or if the test fails constantly/repeatedly, contact Service.
WARNING
Do not treat patients unless the reason for a QA Test failure is clear and it is certain that
there is no increased risk to the patient.

7.2.4. Storing the QA phantom

The properties of the QA phantom affect the test results. If the phantom is damaged the results
may be unreliable. Therefore, follow the handling and storage instructions of the phantom and if
necessary order a new phantom.
The QA Phantom Plate serves as a storage platform for the QA phantom. The QA Phantom Plate
elevates the QA phantom so that the membrane does not rest on any surface. This is to avoid
damaging the sensitive membrane.
1.
Place the QA Phantom Plate on a hard flat level surface. Make sure that within the opening of
the plate there are no objects that could puncture the membrane of the phantom.
2.
Place QA phantom on the QA Phantom Plate with the membrane facing down.
Instructions for Use
109745C2 / 02-2022
1.
Check that the phantom is a sonication
phantom provided by Profound Medical.
2.
Check that the QA phantom is located
correctly on the QA Phantom Plate.
3.
Check that the MR scanner's couch lifting unit
is fully lowered.
4.
Check that there are no sonication holes inside
the QA phantom. If the phantom is damaged,
replace it.
5.
Repeat the QA Test.
Repeat the QA Test.
146 (192)