Error Recovery - Profound Sonalleve MR-HIFU Instructions For Use Manual

Table of Contents

Advertisement

6. Workflow > 6.10. Error recovery

6.10. Error recovery

The system automatically saves the current stage of the treatment as a Saved State file.
WARNING
If the Saved State cannot be loaded, wait until the tissue has cooled before continuing the
treatment. Without the Saved State data, the system cannot estimate the effect of earlier
accumulated heating when recommending cooling times to prevent burns.
If the software crashes, take the following steps to recover:
1.
Restart the Therapy application.
The Save State recovery dialog appears, asking you if you want to load the data from the
2.
previous session. Answer Yes.
If the dialog does not appear automatically, select Manage States from the main menu and
3.
select the last state from the Session Manager.
4.
Perform a test sonication with low power before continuing the therapy after crash recovery.
If the Saved State cannot be loaded, wait until the tissue has cooled before continuing the
treatment. Without the Saved State data, the system cannot estimate the effect of earlier
accumulated heating when recommending cooling times to prevent burns.
Instructions for Use
109745C2 / 02-2022
Fig. 100: Save State recovery dialog.
128 (192)